Gas Receiver Combination Oxygen & Toxic Gas Detector/GTD-6000
Information

Self diagnosis function with built-in Microprocessor
Superior performance for detecting toxic gases below TLV
Easy installation with a wall mount type
4-digit gas concentration display
3-stage alarm level, 1-trouble LED display and relay contact output
Easy configuration of monitoring by RS-485 Modbus communication
Programmable alarm setting (Level, Time delay)
Self diagnosis
4-20mA(Current source or Sink) output signal

No drive file values found.

 

Specification Value
   Measuring gas O2 Toxic gases(CO, Cl2, NH3 & etc)
   Measuring type Diffusion type
   Measuring method Galvanic cell or Electrochemical cell Electrochemical cell
   Measuring range Refer to Toxic Gas Measuring Range
   Response time < 15sec / 90% scale < 45sec / 90% scale
   Accuracy ±3% / Full Range

   Measuring display

1.8″ 4-Digit FND

   Operation mode display

6-LED(Power, 3-stage alarm, Trouble, Stand-by for maintenance)

   

   Alarm signal output

Visual : 3-stage Alarm
Trouble LED
 Warning LED
Audible : Buzzer(85dB)
   

   

  

   

   

   

   Output signal

   

Analog : 4-20mA Analog Output
Maintenance signal : 4mA
Calibration signal : 3mA
Fault signal : 0mA
Digital : RS-485 Modbus(Option)

    Distance

4 – 20mA DC Signal : 2000m RS-485 Modbus Signal : 1000m
   Mounting type Wall mount
   Operating temperature   -20°C ~ +60°C
   Operating humidity  5 to 99% RH (Non-condensing)

   Material 

Detector(ABS), Sensor(PVC), Alarm unit(Poly carbonate)
   Dimensions 235.6(W) × 355.5(H) × 109.3(D) mm

   Weight